DSCUSB High Stability Bridge Measurement | |||||
---|---|---|---|---|---|
Min | Typical | Max | Units | ||
Strain Gauge Excitation System | 4 Wire | ||||
Strain Gauge Excitation Voltage | 5 | V DC | |||
Strain Gauge Drive Capability | 85 | to | 5000 | Ohms | |
Strain Gauge Sensitivity | 3 | mV/V | |||
Offset Temperature Stability | 4 | ppm/C | |||
Gain Temperature Stability | 5 | ppm/C | |||
Offset Stablitity with Time | 90 | ppm of FR (1) | |||
Gain Stability with Time | 30 | ppm of FR (2) | |||
Non Linearity before Linearization | 25 | ppm of FR | |||
Internal Resolution | 16 Million | Counts/Divs | |||
Resolution @ 4.8Hz Readings (Noise Stable) over 1s | 8192 | Counts/Divs | |||
Resolution @ 4.8Hz Readings (Noise Stable) over 1s | 13 | Bits | |||
Measurement Sample Rate | 4800 | Per Second | |||
Measurement Sample Rate Accuracy | ±1.1 % | % | |||
Signal Filter | Dynamic recursive type user programmable | ||||
Electrical | |||||
Supply Voltage Range | 5 | V DC | |||
Average Operational Current (normal mode) | 68 | mA | |||
Data Transmission | |||||
Output Cable Length | 5 | m | |||
Communications (USB) | |||||
The load cell to USB adaptor will communicate as a simple serial device rather than a ‘native’ USB device. Once the device is plugged into a PC and the supplied drivers are installed, the device will appear as a Virtual Serial Port to the PC. |
|||||
Support Modules | |||||
Instrument Explorer Software | |||||
DSC Toolkit Software | |||||
DSCLOG24 Software | |||||
Environmental | |||||
IP Rating | IP50 | ||||
Storage Temperature | - 40 to + 85ºC | ||||
Operating Temperature | - 40 to + 85ºC | ||||
Relative Humidity | 95% maximum non condensing | ||||
Protection | IP50 |
A: FSU modules can connect to a PC by plugging into a USB port and do not require an external power supply as they appear as a ‘single unit load’ i.e. they draw <100mA.
Appropriate USB drivers must be installed which are automatically installed with the FSU Toolkit software.
Ensure the Toolkit is installed before plugging in the FSU device.
A: 4800 sps - this is a fixed rate.
A: By using the free Toolkit software which is available to download from our website. It allows configuration, calibration, data recording and review and for parameter management of the modules.
A: 13 bit resolution, which equates to 1 part in 8000 noise free over a 1 second period.
A: The module indicates status via a red Healthy LED. When the FSU is placed in configuration mode the LED will be on solidly.
When the FSU is in recording mode and transmitting data the LED should flash ON for 100ms every 500ms (LED will be OFF but flash on for a tenth of a second every half a second).
If the integrity of the load cell is in doubt the LED will flash OFF for 100ms every 500ms (LED will be ON but flash off for a tenth of a second every half a second).
A: FSU modules can connect to a PC by plugging into a USB port and do not require an external power supply as they appear as a ‘single unit load’ i.e. they draw <100mA.
A: The free software is downloadable from the Support section of this site
A: The FSU Toolkit is a configuration and analytical tool designed specifically for configuring FSU modules. This toolkit allows configuration, calibration, data recording and review and for parameter management of the modules.
A: Only one (1) FSU at a time can be connected to a PLC.
A: Yes, a DIN rail mounting kit is available for top hat DIN rail. Optional part DSCUSB-DIN consists of a pair of screws and two DIN rail clips.
A: CE Approvals
European EMC Directive, 2004/108/EC
BS EN 61326-1:2006
BS EN 61326-2-3:2006
A: All FSU products from Mantracourt Electronics Ltd., ('Mantracourt') are warranted against defective material and workmanship for a period of three (3) years from the date of dispatch.
70.5 x 51 x 20mm excluding 9-way ‘D’ type socket connector.
74.5 x 51 x 20mm including connector.